ICP Spectrometer LICP-A11 detects micro and macro elements using argon plasma with an analyzing rate of 15 elements per minute. Results can be achieved with lower levels of detection for qualitative and quantitative analysis. Detection of liquid flow from the spray chamber by optical drain sensor offers exceptional sensitivity and accuracy. Accurate signal-background ratio is attained by advanced control system.
Available Range :
Wavelength range ( 3600 lines / mm ) : 180 to 500 nm Wavelength range : 180 to 900 nm Monochromator Specifications : Optical type : Czerny turner Solid State Power Specifications : Frequency : 27.12 MHz| Wavelength range | 180 to 900 nm |
| Elements per minute | 15 elements |
| Temperature | 29 °C |
| Incident slit | 25 µm |
| Standard deviation | RSD ≤ 2 % |
| Correlation coefficient | ≥ 0.9995 % |
| Gas source | Argon ( 99.99 % ) |
| Plasma gas flow meter | 100 – 1000 L / h ( 1.6 - 16 L / min ) |
| Auxillary gas flow meter | 10 – 100 L / h ( 0.16 – 1.66 L / min ) |
| Carrier gas flow meter | 0 ~ 5 L / min |
| Cooling water flow rate | 5 L / min |
| Cooling water temperature | 20 ~ 26 °C |
| Water outlet diameter | 10 mm |
| Line dispersion rate | 0.26 nm |
| Resolving power | ≤ 0.007 nm |
| Co-axial sprayer diameter | 6 mm |
| Coil diameter | 25 mm |
| Double room fog diameter | 34 mm |
| Quartz tube torch diameter | 20 mm |
| Mirror dimensions | 80 x 110 x 16 mm |




Popular Products
Navigation
Contact Us Map