Cytotoxic Safety Cabinet LCSC-A10 comes with Safety Height 200 mm (8 inches) of test opening, 440 mm (17 inches) of maximum opening, 465 m3/h (273 cfm) of exhaust volume and 0.73m3 of internal work area. Maintaining cleanliness within the work zone for patient safety and product cleanliness. Provide highest level of operator safety to protect the user’s personnel from hazardous compounds used in cytotoxic drugs. Incorporated with an advanced range of additional safety features. Spacious knee room is provided for the operator in a sitting position, ensuring comfort during use.
Available Range :
Exhaust Volume : 465 m3/h (273 cfm) Exhaust Volume : 347 m3/h (204 cfm)| Exhaust Volume | 465 m3/h (273 cfm) |
| Tested Opening | Safety Height 200 mm (8 inches) |
| Max Opening | 440 mm (17 inches) |
| Internal Work Area, Space | 0.73 m3 |
| Inflow Velocity | 0.53 m/s |
| Down Flow Velocity | 0.33 m/s |
| Air Supply Filter | ULPA filter, efficiency > 99.999 % for particle size at 0.3 µm |
| Exhaust Filter | 2 pieces HEPA filters, efficiency > 99.999 % for particle size at 0.3 µm Filter life indicator and filter working time indicator |
| Alarm | Abnormal airflow velocity; Filter replacement; Front window at unsafe height |
| Noise | EN 12469 ≤ 58 dB/NSF 49 ≤ 61 dB |
| Water And Gas Tap | Water Tap×1, Gas tap×1 |
| UV Lamp | 30 W×1 Emission of 253.7 nanometres for most efficient decontamination UV life indicator |
| LED lamp | 16 W×2 |
| Illumination | ≥ 1000 lux |
| Consumption | 900 W |
| Waterproof Socket | Two; Total load of two sockets: ≤ 500 W |
| Display | LCD Display |
| Control System | Microprocessor |
| Humidity And Temperature | Humidity and temperature can be displayed on the control panel |
| Material | Work Area:304 stainless steel Main Body: Cold- rolled steel with anti- bacteria powder coating |
| Front Window | Motorized, two-layer laminated toughened glass ≥ 5 mm, anti UV |
| Caster | Foot master Caster |
| Power Supply | AC 220 V ± 10 %, 50/60 Hz; 110 V ± 10 %, 60Hz |
| Internal Size (W×D×H) | 1220×600×630 mm |
| External Size (W×D×H) | 1380×760×2100 mm |
| Package Size (W×D×H) | 1540×910×2280 mm |
| Gross Weight | 371 kg |
| Accessories No. | Accessories Name | Unit |
| 1. | Base Stand | 1 Piece |
| 2. | Drain Valve | 1 Piece |
| 3. | Waterproof Sockets | 2 Pieces |
| 4. | Remote Control | 1 Piece |
| 5. | UV Lamp | 2 Pieces |
| 6. | LED Lamp | 1 Piece |
| 7. | Water Tap | 1 Piece |
| 8. | Gas Tap | 1 Piece |
| Accessories No. | Accessories Name |
| 1. | Armrest |
| 2. | Formalin Fumigation Sterilizer |
| 3. | Infrared Sterilizer |
| 4. | Airflow Tester |
| 5. | PVC Duct |
Labtron offers highly efficient and advanced technology based Class II, Type A2 Biological Safety Cabinets which provides protection to product, personnel and environment against airborne biological or particulate chemical agents. our class II biosafety cabinet storage ensures no leakage and free air return mode in the front suction channel. These products are certified by NSF/ANSI49 and are safe for the users.




Our Vertical Laminar Flow Cabinet provides high degree of contamination free environment. It features tempered glass window for better observation and also provides protection from UV rays. It is widely used in medical, pharmaceutical, electronic and industrial sectors.












Labtron Cytotoxic Safety Cabinet protects users and environment from hazardous cytotoxic agents. They feature effective filtration systems with an exhaust volume range of 207 to 273 CFM. They are equipped with HEPA filters to capture airborne particles and contaminants. They have integrated glove ports for handling cytotoxic materials. Our Cytotoxic Safety Cabinet ensures safe manipulation of cytotoxic agents during diagnostic testing and research.

Labtron Horizontal Laminar Flow Cabinet directs filtered air horizontally across work surface to keep sterile environment. They protect samples from airborne contaminants by maintaining clear airflow. They feature HEPA filters for effective particle removal and smooth easy-to-clean work surface. Our Horizontal Laminar Flow Cabinets are essential for microbiological research, pharmaceutical preparation, and clinical diagnostics.