| Conductivity |
| Conductivity measurement range | Conductivity accuracy | Conductivity calibration points | Conductivity calibration solutions | Conductivity resolution | | 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200 mS/cm | ±0.5% F.S | 1 to 5 points | 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm or Custom | 0.001, 0.01, 0.1, 1 |
|
| TDS |
| TDS range | TDS accuracy | TDS factor | TDS resolution | | 0 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt) | ±1 % F.S | 0.01 to 1.00 (Default 0.5) | 0.01, 0.1, 1 |
|
| Salinity |
| Salinity range | Salinity accuracy | Salinity measurement modes | Salinity resolution | | 0.00 to 80.00 ppt, 0.00 to 42.00 psu | ±1% F.S | Practical Salinity (psu) or Natural Seawater (ppt) | 0.01, 0.1, 1 |
|
| Resistivity |
| Resistivity range | Resistivity accuracy | Resistivity resolution | | 0.00 to 20 MΩ | ±1% F.S | 0.01, 0.1, 1 |
|
| Temperature |
| Temperature measuring range | Temperature measuring accuracy | Temperature calibration points | Temperature resolution | | 0 105 °C, 32 to 221 °F | ± °0.5 C, ± 0.9 °F | 1 point | 0.1 °C, 0.1 °F |
|
| Temperature compensation |
0 to 100 °C, 32 to 212 ° F, Manual or Automatic
|
| Temperature coefficient |
Linear and Pure water compensation
|
| Reference temperature |
20 °C or 25 °C
|
| Stability conditions |
Low or high
|
| Cell constant |
K= 0.1, 1, 10 or custom
|
| Calibration due |
1 to 31 days or off
|
| Memory |
500 data sets
|
| Data hold |
Manual or auto-end point
|
| Connector |
6 pin mini plug
|
| Power Control |
Manual or Automatic (10, 20 or 30 minutes after last key pressed)
|
| Power |
3 × 1.5 V AA batteries or DC 5 V power adapter
|
| Dimension (L × W × H) |
170 × 85 × 30 mm
|
| Weight |
300 g
|